210 Agilent 1260 Infinity DAD and MWD User Manual
10 Maintenance
Replacing Capillaries on a Standard Flow Cell
1
Identify the inlet and outlet capillaries. To replace the inlet
capillary, continue with step "To replace the inlet
capillary, use a 4-mm wrench for the fitting."
2 After replacing the outlet capillary, fix it handtight first.
Then do a 1/4 turn with a 4-mm wrench.
3 To replace the inlet capillary, use a 4-mm wrench for the
fitting.
4 Unscrew the cell body from the heat exchanger and the
heat exchanger from the clamp unit.