Agilent 1260 Infinity DAD and MWD User Manual 165
Test Functions
9
Filter Test
Filter Test Evaluation
Filter Test Failed
Test Failed
Holmium Oxide Maximum out of Limits
Probable cause Suggested actions
1 Filter assembly (lever and filter) not installed.
Install the filter assembly.
2 Defective filter motor.
Exchange the filter motor.
Probable cause Suggested actions
1 Holmium oxide filter not installed.
Install the holmium oxide filter.
2 Dirty or contaminated filter.
Exchange the holmium oxide filter.