226 Agilent 1260 Infinity DAD and MWD User Manual
10 Maintenance
Correcting Leaks
Correcting Leaks
1 Use tissue to dry the leak sensor area and the leak pan.
2 Observe the capillary connections and the flow cell area for leaks and
correct, if required.
Figure 77 Observing for Leaks
3 Replace the front cover.
When If a leakage has occurred in the flow cell area or at the heat exchanger or at the capillary
connections
Tools required Tissue
Two 1/4 inch wrenches for capillary connections
Preparations Remove the front cover.