Agilent 1260 Infinity DAD and MWD User Manual 225
Maintenance
10
Cleaning or Exchanging the Holmium Oxide Filter
4
While releasing the holder with a screw driver, carefully
insert the holmium oxide filter.
5 Replace the flow cell cover and fix the six screws.
Next Steps:
6 Perform a holmium oxide test, see “Holmium Oxide Test” on page 174 to check the proper function of the holmium oxide
filter.
7 Insert the flow cell, see “Exchanging a Flow Cell” on page 200.
8 Replace the front cover.
9 Turn on the flow.