224 Agilent 1260 Infinity DAD and MWD User Manual
10 Maintenance
Cleaning or Exchanging the Holmium Oxide Filter
1
Unscrew the six screws and remove the flow cell cover. 2 If not already in this position, move the filter up.
3 While releasing the holder with a screw driver (at the
top), carefully remove the holmium oxide filter.
NOTE
Do not scratch the holmium oxide filter.
The holmium oxide filter can be cleaned with alcohol
and a lint-free cloth.