Agilent 1260 Infinity DAD and MWD User Manual 149
Error Information
8
General Error Messages
2 If it is desired that the system has only one leak drain tube, then it's
possible in increase the length of the small plastic tube which is connected
to port# 4 of the sampler switching valve. This tube can then be taken to
waste separately. The tube which normally serves as the detector cell outlet
tube can be used for this purpose.
Open Cover
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.
Cover Violation
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
Probable cause Suggested actions
1 The top foam was removed during operation.
Reinstall the top foam.
2 Foam not activating the sensor.
Please contact your Agilent service
representative.
3 Dirty or defective sensor.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 The top foam was removed during operation.
Please contact your Agilent service
representative.
2 Foam not activating the sensor.
Please contact your Agilent service
representative.