Agilent 1260 Infinity DAD and MWD User Manual 155
Error Information
8
Detector Error Messages
Holmium Oxide Test Failed
Illegal Value From Temperature Sensor
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
Probable cause Suggested actions
1 Lamps switched off.
Ensure the lamps are switched on.
2 Defective or dirty flow cell.
Ensure the flow cell is inserted correctly, and is
free from contamination (cell windows, buffers
etc.).
3 Defective filter assembly.
Exchange the filter assembly.
4 Defective achromat assembly.
Exchange the achromat assembly.
5 Defective PDA/optical unit.
Exchange the optical unit.
Probable cause Suggested actions
1 Dirty or defective sensor.
Please contact your Agilent service
representative. Links:
2 Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within the
allowed range.