150 Agilent 1260 Infinity DAD and MWD User Manual
8 Error Information
Detector Error Messages
Detector Error Messages
These errors are detector specific.
Visible Lamp Current
The visible lamp current is missing.
The processor continually monitors the lamp current during operation. If the
current falls below the lower current limit, the error message is generated.
Probable cause Suggested actions
1 Lamp disconnected.
Ensure the visible lamp connector is seated
firmly.
2 Defective visible lamp.
Exchange the visible lamp.
3 Defective connector or cable.
Please contact your Agilent service
representative.
4 Defective power supply.
Please contact your Agilent service
representative.