156 Agilent 1260 Infinity DAD and MWD User Manual
8 Error Information
Detector Error Messages
Illegal Value From Air Inlet Temperature Sensor
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Wavelength Recalibration Lost
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
Probable cause Suggested actions
1 The temperature sensor is defect.
Please contact your Agilent service
representative.
2 Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within the
allowed range.
Probable cause Suggested actions
1 The detector is new.
Recalibrate the detector.
2 The detector has been repaired.
Please contact your Agilent service
representative.

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals