86 Agilent 1260 Infinity DAD and MWD User Manual
5 Using the Detector
Setting up an Analysis
Requirements and Conditions
Parts and Material required
Table 21 on page 86 lists the parts and material you need for the set up of the
analysis. Some of these are optional (not required for the basic system).
Tab le 21 Parts and Material required
Agilent 1260
Infinity system
Pump (plus degassing)
Autosampler
Detector, standard flow cell installed
Agilent ChemStation or
Instant Pilot G4208 (optional for basic operation) or
with with the appropriate revisions, see “Performance Specifications” on
page 38.
System should be correctly set up for LAN communication with the Data
System
Column: Zorbax Eclipse XDB C18, 150 mm x 4.6 mm, 5 µm (p/n 993967-906)
Standard: Agilent isocratic checkout sample (p/n 01080-68704). This 0.5 mL
ampoule contains 0.15 wt.% dimethylphthalate, 0.15 wt.% diethylphthalate,
0.01 wt.% biphenyl, 0.03 wt.% o-terphenyl in methanol.