Agilent 1260 Infinity DAD and MWD User Manual 151
Error Information
8
Detector Error Messages
Visible Lamp Voltage
The visible lamp voltage is missing.
The processor continually monitors the voltage across the lamp during
operation. If the lamp voltage falls below the lower limit, the error message is
generated.
Diode Current Leakage
When the detector is switched on, the processor checks the leakage current of
each of the optical diodes. If the leakage current exceeds the upper limit, the
error message is generated.
Probable cause Suggested actions
1 Defective connector or cable.
Exchange the detector main board.
2 Defective power supply.
Exchange the power supply.
Probable cause Suggested actions
1 Defective PDA/optical unit.
Please contact your Agilent service
representative.
2 Defective connector or cable.
Please contact your Agilent service
representative.