152 Agilent 1260 Infinity DAD and MWD User Manual
8 Error Information
Detector Error Messages
UV Lamp Current
The UV lamp current is missing.
The processor continually monitors the anode current drawn by the lamp
during operation. If the anode current falls below the lower current limit, the
error message is generated.
UV Lamp Voltage
The UV lamp anode voltage is missing.
The processor continually monitors the anode voltage across the lamp during
operation. If the anode voltage falls below the lower limit, the error message is
generated.
Probable cause Suggested actions
1 Lamp disconnected.
Ensure the UV lamp connector is seated firmly.
2 Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
3 Defective detector main board.
Please contact your Agilent service
representative.
4 Defective power supply.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
2 Defective detector main board.
Please contact your Agilent service
representative.
3 Defective power supply.
Please contact your Agilent service
representative.

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals