162 Agilent 1260 Infinity DAD and MWD User Manual
9 Test Functions
Self-test
Self-test
The DAD self-test (see Figure 64 on page 163) runs a series of individual tests,
and evaluates the results automatically. The following tests are run:
Filter Test
Slit Test
Dark Current Test
Intensity Test
Wavelength Calibration Test
Holmium Test
Spectral Flatness Test
ASTM Noise Test (optional)
The self-test can be run once or repetitively. When set up to run repetitively,
the tests run in series continually until stopped by the user. Running the test
repetitively is useful when troubleshooting problems which occur
intermittently.
The ASTM noise test determines the detector baseline noise (254nm) while
pumping water at 1ml/min. The test requires approximately 20 minutes to
run, and can be included or excluded from the self-test sequence as required.
Setup of the self test is done in the Self Test dialog box. Select either Single
Test or Repetitive Tests. Check the ASTM Noise Test checkbox to include the
noise test in the self test.

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals