186 Agilent 1260 Infinity DAD and MWD User Manual
9 Test Functions
Diagnosis Information on Agilent ChemStation
Diagnosis Information on Agilent ChemStation
The Agilent ChemStation provides certain information on various assemblies
in the detector.
1 Open the Diagnosis and change to detailed view.
2 Click on the assembly of interest and select “Update Variables Display“.
Figure on page 187 shows an example.
Table 33 on page 186 lists the assemblies which provide detailed information.
Tab le 33 Diagnosis - detailed information
Assembly Details
General Product number, firmware revision, manufacturing date (of
intsrument or, if replaced, of main board), serial number,
accumulated on-time, spectrometer serial number.
Vis lamp Accumulated on-time, actual on-time, lamp switch on,
accumulated on time switches, switch on voltage/current, lamp
current, lamp voltage/current
UV lamp Accumulated on-time, actual on-time, lamp ignitions,
accumulated on time reset, ignition voltage/current, lamp
voltage, lamp voltage/current.
Available with ID-tag lamp only.
product number, serial number, production date, last intensity
test
Holmium Filter Filter movements, filter moves reset log
Flow Cell Available with ID-tag flow cells only.
Product number, production date, pathlength, max pressure,
serial number, volume, last cell test
Micro Slit Slit movements, filter moves reset log