Agilent 1260 Infinity DAD and MWD User Manual 211
Maintenance
10
Replacing Capillaries on a Standard Flow Cell
5
Use a small flat screw driver to carefully lift off the I.D.
tag. Shown is the default orientation. See Note at the
beginning of this section.
6 Unscrew the fixing screw and unwrap the inlet capillary
from the grove in the flow cell body.
7 Take the new inlet capillary and bend it 90° about 35 mm
from its end.
8 Bend the capillary again by 90° as shown below.