212 Agilent 1260 Infinity DAD and MWD User Manual
10 Maintenance
Replacing Capillaries on a Standard Flow Cell
9
Insert the capillary into the hole between fixing screw and
the inlet fitting..
10 The capillary lays in the grove and should be tied around
the body (in the grove) 5 times.
11 Insert the fixing screw, so that the capillary cannot leave
the grove.
12 Carefully insert the I.D. tag into the new heat exchanger.
Shown is the default orientation. See Note at the
beginning of this section.