© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-86
Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000)
Table 10-9 Series 3 and 79000 family power requirements (single-phase power)
E1135C
Power
Option Voltage Full-Load Amps (FLA)
1
for:
2
327X / 79000
3
317X 307X xDSL/POTS
Bay
4
Opts/Current
Testhead Support Bay
Single-Phase Wye w/
Neutral
AWV 120 /
208
[25]
5
20 [43] 24 [42] 28 [24] 15 AWY, 16
AWZ 127 /
220
[25] 20 [43] 24 [42] 28 [24] 15 AWZ, 16
05B 220 /
380
[18] 11 [21] 12 [37] 22 NA 05M, 14
05G 230 /
400
[18] 11 [21] 12 [37] 22 NA 05F, 14
05K 240 /
415
[18] 11 [21] 12 [37] 22 NA 05J, 14
Single-Phase Earthed 05M 220 [22] 20 [36] 24 [42] 28 [24] 15 14
05F 230 [22] 20 [36] 24 [42] 28 [24] 15 14
05J 240 [22] 20 [36] 24 [42] 28 [24] 15 14
L1 L2
G
N
L1
N
G