© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-87
Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000)
327X / 79000 317X 307X xDSL/POTS
Bay
Opts/Current
Single-Phase
Non-Earthed
AWX 200 [NA] 20 [36] 24 [42] 28 [20] 15 AWY, 16
0EB 220 [NA] 20 [36] 24 [42] 28 [20] 15 0E3, 14
0EJ 230 [NA] 20 [36] 24 [42] 28 [20] 15 0EG, 14
0EC 240 [NA] 20 [36] 24 [42] 28 [20] 15 0EH, 14
Single-Phase Center-Tap
Neutral
AWY 100 /
200
[25] 20 [43] 24 [43] 28 [24] 15 16
0E3 110 /
220
[25] 20 [43] 24 [43] 28 [24] 15 14
0EG 115 /
230
[25] 20 [43] 24 [42] 28 [24] 15 14
0EH 120 /
240
[25] 20 [43] 24 [42] 28 [24] 15 14
1. Calculated with a fully-loaded 3X7X system with 3 STC cards in every module.
2. DUT Power supplies are configured for 200-240 volts ac.
Table 10-9 Series 3 and 79000 family power requirements (single-phase power) (continued)
E1135C
Power
Option Voltage Full-Load Amps (FLA)
1
for:
2
L1
L2
G
L1
L2
G