Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II)
Single-Phase
Non-Earthed
AWX 200 Table 10-41 on page 10-142
0EB 220 Table 10-41 on page 10-142
0EJ 230 Table 10-41 on page 10-142
0EC 240 Table 10-41 on page 10-142
Single-Phase Center-Tap
Neutral
AWY 100 / 200 Table 10-39 on page 10-141
0E3 110 / 220 Table 10-39 on page 10-141
0EG 115 / 230 Table 10-39 on page 10-141
0EH 120 / 240 Table 10-39 on page 10-141
Table 10-34 Conductor and breaker size cross reference table (single-phase options) (continued)
System Power
Option Voltage Wire and Breaker Size Table Number
L1
L2
G
L1
L2
G
N