© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-139
Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II)
Table 10-35 3-phase wye with neutral (0EF, 05E)
System Type
POTS Bay 327X 317X 307X
Breaker Size in Amps
20 20 30 50
SO Cord Conductor Size mm2 (American Wire Gauge - AWG)
4 (12) 6 (10) 6 (10) 10 (8)
Table 10-36 3-phase wye with neutral (0E5, 05L, 0E7)
System Type
327X 317X 307X
Breaker Size in Amps
20 20 30
SO Cord Conductor Size mm2 (American Wire Gauge - AWG)
4 (12) 4 (12) 6 (10)