EasyManuals Logo

Agilent Technologies 41501B User Manual

Agilent Technologies 41501B
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #212 background imageLoading...
Page #212 background image
8-4 Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
Specifications
Hardware
Hardware
4156C Precision Semiconductor Parameter Analyzer
HRSMU (High Resolution SMU) Specifications
Current Range, Resolution, and Accuracy: See Table 8-1.
Voltage Range, Resolution, and Accuracy: See Table 8-3.
Output terminal/connection
Dual triaxial connectors, Kelvin (remote sensing)
Voltage/Current Compliance (Limiting)
The SMU can limit output voltage or current to prevent damaging the device
under test.
Voltage 0 V to ± 100 V
Current ± 100 fA to ± 100 mA
Compliance accuracy Same as current (voltage) set accuracy.
HRSMU supplemental information
Allowable cable resistance maximum 10 Ω (for both Force and Sence. Kelvin
connection.)
Voltage source output
resistance typical 0.2 Ω (Force line. Non-Kelvin connection.)
Voltage measurement input
resistance 10
15
Ω
Current source output
resistance 10
15
Ω (10 pA range)
Current compliance setting
accuracy (for opposite polarity)
For10pAto10nArange:
V/I setting accuracy ± 12 % of range
For 100 nA to 100 mA range:
V/I setting accuracy ± 2.5 % of range

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 41501B and is the answer not in the manual?

Agilent Technologies 41501B Specifications

General IconGeneral
BrandAgilent Technologies
Model41501B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals