Testing Performance 2
5000 Series Oscilloscopes Service Guide 37
8 Make the following measurements.
a For 300 MHz and 500 MHz models, period 5 ns — the test
limits are 4.96 ns to 5.04 ns.
b For 100 MHz models, period 10 ns — the test limits are
9.93 ns to 10.07 ns.
If the measurements are not within the test limits, go to the
“Troubleshooting” chapter. Then return here.
9 Record the results as Pass or Fail in the Performance Test
Record (see page 48).