38 5000 Series Oscilloscopes Service Guide
2 Testing Performance
To verify trigger sensitivity
This test verifies the trigger sensitivity. In this test, you will
apply a sine wave to the oscilloscope at the upper bandwidth
limit. You will then decrease the amplitude of the signal to the
specified levels, and check to see if the oscilloscope is still
triggered.
The internal trigger sensitivity test is mandatory because it is a
specification. The external trigger test is optional because it is a
characteristic, not a specification.
Test limits for:
Internal trigger on all models:
< 10 mV/div: greater of 1 div or 5 mV
pp
>=10 mV/div: 0.6 div
External trigger on all 2-channel models (DSO5xx2A):
Trigger range: 1.0V
DC to 100 MHz: < 100 mV
pp
>100 MHz to max bandwidth: < 200 mV
pp
Trigger range: 8.0V
DC to 100 MHz: < 250 mV
pp
>100 MHz to max bandwidth: < 500 mV
pp
External trigger on 4-channel models:
4-channel models:
DC to 100 MHz: < 500 mV
pp

Table of Contents

Other manuals for Agilent Technologies 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5000 Series and is the answer not in the manual?

Agilent Technologies 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals