74 5000 Series Oscilloscopes Service Guide
4 Troubleshooting
To verify default setup
The oscilloscope is designed to turn on with the setup from the
last turn on or previous setup. However, if the Secure
Environment option is installed, the oscilloscope will always
execute a Default Setup upon power-up.
To recall the factory default setup:
1 Press the Save/Recall key.
2 Press the Default Setup softkey.
This returns the oscilloscope to its factory default settings
and places the oscilloscope in a known operating condition.
The major default settings are:
Horizontal - main mode, 100 us/div scale, 0 s delay, center
time reference
Vertical - Channel 1 on, 5 V/div scale, dc coupling, 0 V
position, probe factor to 1.0 if an AutoProbe probe is not
connected to the channel
Trigger - Edge trigger, Auto sweep mode, 0 V level, channel
1 source, dc coupling, rising edge slope, 60 ns holdoff time
Display - Vectors on, 20% grid intensity, infinite persistence
off
Other - Acquire mode normal, Run/Stop to Run, cursor
measurements off

Table of Contents

Other manuals for Agilent Technologies 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5000 Series and is the answer not in the manual?

Agilent Technologies 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals