Testing Performance 2
5000 Series Oscilloscopes Service Guide 39
Tab le 8 Equipment Required to Verify Trigger Sensitivity
Equipment Critical Specifications
Recommended
Model/Part Qty
Signal Generator 100 kHz - 500 MHz at 200 mVrms Agilent N5181A, Agilent
E4400B, or Agilent 8648A
1
Power Splitter Outputs differ 0.15 dB Agilent 11667A 1
Cable BNC Agilent 10503A 1
Adapter N (m) to BNC (f) Agilent 1250-0780 1
Feedthrough 50 BNC (f) to BNC (m) Agilent 0960-0301 1