Replacing Assemblies 5
5000 Series Oscilloscopes Service Guide 93
Figure 25 Removing the display assembly
6 To reinstall the display:
a Align the top locating tab with the locating slot in the
sheet metal and the retaining hooks with their retaining
holes.
b Push down on the LCD until it is flat with the deck and at
the same time push the LCD to your left.
c Ensure the latch tab is seated in its hole as shown in
Figure 25.
d Route the cables down through the deck hole and
reconnect to the system board as shown in Figure 24.
Display (LCD)
Assembly
Latch Slot
Latch Tab
Deck
Retaining Hooks