Calibrating and Adjusting
User Calibration 53
This chapter explains how to adjust the oscilloscope for 
optimum operating performance. You should perform 
self-calibration according to the following recommendations:
• Every 12 months or after 2000 hours of operation
• If the ambient temperature is >10 °C from the calibration 
temperature
• If you want to maximize the measurement accuracy
The amount of use, environmental conditions, and experience 
with other instruments help determine if you need shorter 
adjustment intervals.
Let the Equipment Warm Up Before Adjusting
Before you start the adjustments, let the oscilloscope and test 
equipment warm up for at least 30 minutes.
Read All Cautions and Warnings
Read the following cautions and warning before making 
adjustments or performing self-calibration.
WARNING
HAZARDOUS VOLTAGES!
Read the safety notice at the front of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with the 
protective covers removed. Only trained service personnel who are aware of the 
hazards involved should perform the maintenance. Whenever possible, perform the 
procedures with the power cord removed from the oscilloscope.