Calibrating and Adjusting 3
5000 Series Oscilloscopes Service Guide 53
User Calibration
Perform user-calibration:
Each year or after 2000 hours of operation.
If the ambient temperature is >10° C from the calibration
temperature.
If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience
with other instruments help determine if you need shorter User
Cal intervals.
User Cal performs an internal self-alignment routine to
optimize the signal path in the oscilloscope. The routine uses
internally generated signals to optimize circuits that affect
channel sensitivity, offset, and trigger parameters. Disconnect
all inputs and allow the oscilloscope to warm up before
performing this procedure.
Performing User Cal will invalidate your Certificate of
Calibration. If NIST (National Institute of Standards and
Technology) traceability is required perform the procedures in
Chapter 2 in this book using traceable sources.
To perform User Cal
1 Set the rear-panel CALIBRATION switch to UNPROTECTED.
2 Connect short (12 inch maximum) equal length cables to
each analog channel’s BNC connector on the front of the
oscilloscope. You will need two equal-length cables for a
2-channel oscilloscope or four equal-length cables for a
4-channel oscilloscope.

Table of Contents

Other manuals for Agilent Technologies 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5000 Series and is the answer not in the manual?

Agilent Technologies 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals