Testing Performance 2
5000 Series Oscilloscopes Service Guide 41
2 Verify the trigger sensitivity at maximum bandwidth.
500 MHz models: 500 MHz
300 MHz models: 300 MHz
100 MHz models: 100 MHz
a Press the Save/Recall key, then press the Default Setup
softkey.
b Set channel 1 Imped to 50 Ohm.
c Set the output frequency of the signal generator to the
maximum bandwidth of the oscilloscope and set the
amplitude to about 10 mV
pp
.
d Press the AutoScale key.
e Set the time base to 10 ns/div.
f Set channel 1 to 5 mV/div.
g Decrease the amplitude from the signal generator until 1
vertical division of the signal is displayed.
The trigger is stable when the displayed waveform is
stable. If the trigger is not stable, try adjusting the trigger
level. If adjusting the trigger level makes the trigger stable,
the test still passes. If adjusting the trigger does not help,
see the “Troubleshooting” chapter. Then return here.
h Record the result as Pass or Fail in the Performance Test
Record (see page 48).
3 Repeat this procedure for the remaining oscilloscope
channels.

Table of Contents

Other manuals for Agilent Technologies 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5000 Series and is the answer not in the manual?

Agilent Technologies 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals