44 5000 Series Oscilloscopes Service Guide
2 Testing Performance
 
2 Set up the oscilloscope.
a Press the Save/Recall key, then press the Default Setup 
softkey.
b Set the External Trigger impedance to 50 Ohm.
c Change the trigger Mode from Auto to Normal.
d Use the Range softkey and the Entry knob to set the range 
to 1.0 V.
3 Verify the trigger sensitivity at maximum frequency.
a Change the signal generator output frequency:
DSO5052A: 500 MHz
DSO5032A: 300 MHz
DSO5012A: 100 MHz
b Set the power meter Cal Factor % to the appropriate value 
(100, 300, or 500 MHz) on the calibration chart on the 
power sensor. If necessary, do a linear interpolation if the 
correct factor is not included in the power meter’s 
calibration chart.
c Adjust the signal generator output for a reading on the 
power meter of 100µW. (200 mV
pp
 = 70.71mV rms, Power = 
Vin
2
/50Ω = 70.71 mV
2
/50Ω = 100µW.)
d Press the Trigger Edge key, then press the Source softkey to 
set the trigger source to external trigger.
e Check for stable triggering and adjust the trigger level if 
necessary. Triggering is indicated by the Trig’d indicator at 
the top of the display. When it is flashing, the oscilloscope 
is not triggered. When it is not flashing, the oscilloscope is 
triggered.
f Record the results as Pass or Fail in the Performance Test 
Record (see page 48).
If the test fails, see the “Troubleshooting” chapter. Then 
return here.