Testing Performance 2
5000 Series Oscilloscopes Service Guide 47
2 Set up the oscilloscope.
a Press the Save/Recall key, then press the Default Setup
softkey.
3 Set the signal generator output frequency to 100 MHz.
4 Set the power meter Cal Factor % to the 100 MHz value on the
calibration chart on the power sensor.
5 Adjust the signal generator output for reading on the power
meter of 625µW. (500mV
pp
= 176.78mV rms, Power =
Vin
2
/50 = 176.78 mV
2
/50 = 625µW.)
6 Press the Trigger Edge key, then press the Source softkey to
set the trigger source to External.
7 Check for stable triggering and adjust the trigger level if
necessary. Triggering is indicated by the Trig’d indicator at
the top of the display. When it is flashing, the oscilloscope is
not triggered. When it is not flashing, the oscilloscope is
triggered.
8 Record the results as Pass or Fail in the Performance Test
Record (see page 48).
If the test fails, see the “Troubleshooting” chapter. Then
return here.

Table of Contents

Other manuals for Agilent Technologies 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5000 Series and is the answer not in the manual?

Agilent Technologies 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals