Chapter 4 Using the Agilent 5530 for the First Time
Task 3: Aligning the Optics
4-10 Getting Started Guide
4
Task 3: Aligning the Optics
In this procedure, you make sure that the optics are aligned with the laser
beam path and with each other. Follow these steps:
1 Set the turret ring on the laser head to OTHER (Figure 4-7).
Figure 4-7. Front of the laser head
2 Rotate the laser heads target into position over the lower port by turning
the lower aperture control.
3 Pass your finger between the interferometer assembly and retroreflector.
One of the alignment dots on the laser head should flicker. This is the
retroreflector return beam. The other is the interferometer assembly
return beam.
NOTE
You may need to adjust the height of the interferometer and or the
reflector to see both return beams on the target in the lower port on the
laser head.
1
Turret ring set to “Other”
2 Upper aperture control
3 Upper port
4 Lower port
5 Lower aperture control
STRAIGHT
AVOID EXPOSURE
LASER RADIATION IS
EMITTED FROM THIS APERTURE
LASER ON SIGNAL
OTHER
4
5
3
2
1
Agilent

Table of Contents

Other manuals for Agilent Technologies 5530

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5530 and is the answer not in the manual?

Agilent Technologies 5530 Specifications

General IconGeneral
BrandAgilent Technologies
Model5530
CategoryTest Equipment
LanguageEnglish

Related product manuals