Configuration 10
Operation Manual 159
The flows to the three detectors are based on the pressure
drops through the capillaries and their resistance to flow. An
Agilent flow calculator provided with the capillary flow
splitter device is used to size the length and diameter of
these capillary sections to obtain the desired split ratios.
Your analytical method can set the flow or pressure for
column # 2, the lowest numbered column in the split. Use
the value obtained from the Agilent flow calculator for this
setpoint in your method.