EasyManuals Logo

Agilent Technologies 7890B User Manual

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
Chromatographic Checkout 6
Operation Manual 59
Septum purge 15 mL/min
PTV inlet
Mode Splitless
Inlet temperature 75 °C
Initial time 0.1 min
Rate 1 720 °C/min
Final temp 1 350 °C
Final time 1 2 min
Rate 2 100 °C/min
Final temp 2 250 °C
Final time 2 0 min
Purge time 0.5 min
Purge flow 40 mL/min
Septum purge 3 mL/min
Detector
Temperature 300 °C
H2 flow 30 mL/min
Air flow 400 mL/min
Makeup flow (N2) 25 mL/min
Lit offset Typically 2 pA
Oven
Initial temp 75 °C
Initial time 0.5 min
Rate 1 20 °C/min
Final temp 190 °C
Final time 0 min
ALS settings (if installed)
Sample washes 2
Sample pumps 6
Sample wash volume 8
Injection volume 1 µL
Syringe size 10 µL
Table 4 FID Checkout Conditions (continued)

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals