126 Chapter3
Manual Adjustment Procedures
4. Sampling Oscillator Adjustment
4. Sampling Oscillator Adjustment
Assembly Adjusted
A15 RF assembly
Related Performance Test
There is no related performance test for this adjustment procedure.
Description
The sampling oscillator tank circuit is adjusted for a tuning voltage of
5.05 V dc when the sampling oscillator is set to 297.222 MHz. The
voltage monitored is actually the tuning voltage divided by 4.05. The
setting is then checked at other frequencies for the full tuning range of
the sampling oscillator.
Figure 3-11 Sampler Adjustment Setup
Equipment
Digital voltmeter ...............................................................3456A
DVM test leads ................................................................34118A