604 Chapter12
RF Section
A8 Low Band Mixer
A8 Low Band Mixer
1. Connect the spectrum analyzer CAL OUTPUT to the INPUT 50
connector.
2. Set the spectrum analyzer as follows:
Center frequency..................................................... 300MHz
Span .................................................................................0Hz
Input attenuation ..........................................................10dB
3. If the spectrum analyzer serial number prefix is 3641A or greater,
make sure A8 is receiving the 5 V and 4 V supply voltages from
frequency control board assembly A14 via cable assembly W12.
4. Using another spectrum analyzer, check for approximately 21 dBm
(300 MHz) at the input of A8. (This level can easily be measured at
the output of FL1 by disconnecting W45 from FL1.)
5. If the level at the input of A8 is less than 25 dBm, suspect FL1
low-pass filter, A10 RYTHM, A12 SBTX, or A9 input attenuator.
Refer to power levels shown on Figure 12-11 on page 631, RF
Section Troubleshooting Block Diagram.
6. Check for approximately 30 dBm (3.9107 GHz) at the output of A8.
(This level can easily be measured at the output of FL2 by
disconnecting W57 from FL2.)
7. If the level at the output of A8 is less than 35 dBm, suspect A8 low
band mixer or FL2 low-pass filter.

Table of Contents

Other manuals for Agilent Technologies 8565EC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8565EC and is the answer not in the manual?

Agilent Technologies 8565EC Specifications

General IconGeneral
BrandAgilent Technologies
Model8565EC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals