72 Chapter2
Adjustment/Diagnostic Software
6. Cal Out Adjustment
6. Cal Out Adjustment
Assembly Adjusted
A15 RF assembly
Related Performance Test
Calibrator Amplitude and Frequency Accuracy
Procedure
1. Connect the 8482A power sensor (through adapter) to the front
panel CAL OUTPUT on the spectrum analyzer.
2. Carefully follow the instructions issued by the software program.
3. If the spectrum analyzer has Option 103, TCXO frequency reference,
the TCXO will also be adjusted.