Chapter 11 577
Synthesizer Section
Sampler and Sampler IF
Sampler and Sampler IF
The A15U100 sampler creates and mixes harmonics of the sampling
oscillator with the 1st LO. The resulting sampler IF (60 MHz to
96 MHz) is used to phase-lock the YTO. The sampler IF filters
unwanted products from the output of A15U100 and amplifies the IF to
a level sufficient to drive the YTO loop. When the IF is between 78 and
87 MHz, PIN diodes switch a 120 MHz notch filter in the sampler IF
section.
1. Set the spectrum analyzer to the following settings:
Center frequency ..................................................... 300MHz
Span .................................................................................0Hz
2. Disconnect W32 from A15J101.
3. Connect the input of a power splitter to A15J101. Connect W32 to
one of the splitter outputs. Connect the other splitter output to the
input of another spectrum analyzer.
4. If a 66.7 MHz signal, greater than 15 dBm, is not displayed on the
other spectrum analyzer, set a microwave source to the following
settings:
Frequency ............................................................ 4.2107GHz
Amplitude .................................................................. 5dBm
5. Connect the microwave source to A15U100J1. A 66.7 MHz signal at
approximately 0 dBm should be displayed on the other spectrum
analyzer.
6. Use an active probe/spectrum analyzer combination to measure the
signal at the following test points:
7. If a correct signal is seen at A15TP201 but the signal at A15TP101 is
wrong, proceed as follows:
Use an oscilloscope to measure the signals at the following test
points:
A15J400 pin 1 +0.8 Vdc to +1.6 Vdc (0.5 Vp-p variation)
A15J400 pin 3 0.8 Vdc to 1.6 Vdc (0.5 Vp-p variation)
If these levels are wrong, perform the "Power and Sampler Match
Adjustments" in the sampling oscillator adjustment procedure.
Refer to Chapters 2 and 3.
A15TP101 66.7 MHz, 8 dBm
A15TP201 296 MHz, +9 dBm

Table of Contents

Other manuals for Agilent Technologies 8565EC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8565EC and is the answer not in the manual?

Agilent Technologies 8565EC Specifications

General IconGeneral
BrandAgilent Technologies
Model8565EC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals