Chapter 4 177
Assembly Replacement
Access to Internal Assemblies
Access to Internal Assemblies
Servicing the 8564EC or the 8565EC requires the removal of the
spectrum analyzer cover assembly and the folding down of six board
assemblies. Four of these assemblies lay flat along the top of the
spectrum analyzer and two lay flat along the bottom of the spectrum
analyzer. All six assemblies are attached to the spectrum analyzer right
side frame using hinges and fold out of the spectrum analyzer allowing
access to all major assemblies. See Figure 4-1 on page 179.
•
To remove the spectrum analyzer cover assembly, refer to “Procedure 1.
Spectrum Analyzer Cover.”
• To access the A2, A3, A4, and A5 assemblies, refer to “Procedure 5.
A2, A3, A4, and A5 Assemblies.”
•To access the A14 and A15 assemblies, refer to“Procedure 8. A14
and A15 Assemblies.”
• To remove and replace the backlight cables, which illuminate the
A18 LCD, refer to“Procedure 2. A1 Front Frame/A18 LCD."
•To remove the A17 LCD board, refer to“Procedure 2. A1 Front
Frame/A18 LCD.”