Manual Adjustment Procedures: 3335A Source not Available
3a. DC Log Amplifier Adjustments
10.If the signal is higher than expected (delta marker amplitude reads
greater than 38 dB) then adjust A4R544 for an even higher level
signal and press
CAL and ADJ CURR IF STATE. Allow sufficient time
for the analyzer to complete the adjustment.
11.Repeat step 7 through step 10.
A4 LOG Fidelity Adjustment
1. Press LINE to turn the spectrum analyzer off. Remove the spectrum
analyzer cover and place the spectrum analyzer in the service
position as illustrated in Figure 3a-3. See Figure 3a-4 for adjustment
location.
2. Connect the E4421B RF output to the spectrum analyzer 50 input.
Press
LINE to turn the spectrum analyzer on.
3. On the spectrum analyzer, press
PRESET, CAL, IF ADJ ON OFF (OFF),
and
ADJ CURR IF STATE.
4. Set the spectrum analyzer controls as follows:
Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 MHz
Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 kHz
Reference level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dBm
5. Set up an E4421B as follows:
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dBm
Mod On/Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off
6. On the spectrum analyzer, press
MKR and MARKER DELTA.
7. Decrease the E4421B power to 26 dBm.
8. Calculate the error:
9. If the error is less than ±0.2 dB, no adjustment is necessary.
10.Set the E4421B power to 10 dBm.
11.Adjust A4R531 (see Figure 3a-4) to read two times the error. For
example, if the calculated error is +0.75 dB, adjust A4R531 for a
delta marker amplitude reading of +1.5 dB. Press
CAL and
ADJ CURR IF STATE.
12.Repeat step 7 through step 11.
Error delta marker reading 16 dB=

Table of Contents

Other manuals for Agilent Technologies 8565EC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8565EC and is the answer not in the manual?

Agilent Technologies 8565EC Specifications

General IconGeneral
BrandAgilent Technologies
Model8565EC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals