168 Chapter3
Manual Adjustment Procedures: 3335A Source not Available
2a. IF Amplitude Adjustments
6. On the spectrum analyzer, reconnect W29 to A5J3. Press PRESET
and set the controls as follows:
Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 Hz
Reference level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 dBm
Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . .300 kHz
7. Connect a BNC cable between the 8563E CAL OUTPUT and
INPUT 50 .
8. On the spectrum analyzer, press
MKR CAL and REF LVL ADJ.
9. Use the knob or step keys to adjust the REF LEVEL CAL setting
until the MKR reads 10.00 dBm ±0.1 dB.
10.On the spectrum analyzer, press
STORE REF LVL.