52 Chapter2
Adjustment/Diagnostic Software
Introduction
A7 LOMALOMA adjustments4
Front end cal (or perform the frequency response
performance test in the 8560 E-Series and
EC-Series Spectrum Analyzer Calibration
Guide. The adjustment must be performed if
the performance test fails).
7
A8 low band
mixer
Front end cal7
A9 input
attenuator
Front end cal (or perform the frequency response
performancetest in the8560 E-Series Spectrum
Analyzer Calibration Guide. The adjustment
must be performed if the performance test
fails).
7
A10/A12Front end cal7
RYTHM/SB
TX
A11 YTO
YTO adjustment
‡
5
LO frequency
‡
2
YTO FM coil
‡
3
A13 2nd
converter
Front end cal7
A14
frequency
control
YTO adjustment
‡
5
LO frequency
‡
2
YTO FM cal
‡
3
Front end cal7
*If any automated adjustment is required, you must first perform automated
adjustment ”1. Initial Information”
†
Either the manual or the automated adjustment may be performed.
Table 2-1 Related Adjustments
Assembly
Changed
or
Repaired
Perform the following related
adjustments in the order listed
Adjustment Number
Manual Automated*