2-32 Service Guide
System Verification and Performance Tests 8719ET/20ET/22ET
System Verification 8719ES/20ES/22ES
RF Network Analyzers
Figure 2-16 Printout of Tabular Verification Results
The callouts in Figure 2-16 are defined as:
1. Frequency of the data points, in GHz.
2. Results of the measurement.
3. Upper limit line as defined by the total system uncertainty specification.
4. Lower limit line as defined by the total system uncertainty specification.
5. Difference between the measured results and the limit line. A positive number indicates
a pass. An asterisk (*) indicates a fail.