Details routine tasks for maintaining the 7890 Series GC, assuming basic tool and GC operation knowledge.
Lists common tools and materials needed for most GC maintenance procedures, with specific tools listed in procedure steps.
Recommends creating and storing GC maintenance methods to prevent damage and injury during maintenance.
Advises reading important safety and regulatory information found in the 7890 Series Safety and Regulatory Information book.
Introduces the Agilent Parts Finder tool for locating replacement and consumable part numbers.
Provides instructions for removing the detector top cover, which protects detectors, valve box, and valve assembly.
Details how to remove the pneumatics cover, which protects flow manifolds in the back top of the GC.
Explains how to remove the electronics cover, noting steps differ based on FPD installation.
Lists consumable parts for the hydrogen sensor module accessory and general GC maintenance.
Illustrates and identifies components for the hydrogen sensor calibration gas setup.
Provides step-by-step instructions for replacing the calibration gas cylinder for the hydrogen sensor.
Explains how to remove and install the GC's removable oven door for easier access.
Lists nuts, ferrules, and hardware for capillary columns, with typical use and part numbers.
Provides instructions for selecting a hanger position and inserting the hanger into the GC oven slots.
Details the process of installing capillary column clips using mounting screws and clips.
Guides on conditioning a capillary column, including gathering materials and setting up the GC.
Explains how to cut a loop from a column, including loosening the nut and cutting the unwanted loop.
Describes how to reverse a column and bakeout contaminants, including disconnecting and reattaching the column.
Details the procedure for attaching a capillary column using SilTite metal fittings and a swaging tool.
Provides instructions on how to disconnect fused silica tubing from a SilTite fitting.
Lists various liners, nuts, ferrules, and other consumables for the split/splitless inlet.
Diagram showing exploded views of the components of the split/splitless inlet.
Step-by-step guide for installing a capillary column into the split/splitless inlet.
Instructions for replacing the septum in the split/splitless inlet, including safety precautions.
Procedure for cleaning the septum seat in the insert assembly of the split/splitless inlet.
Details on how to change the liner and O-ring in the split/splitless inlet.
Instructions for replacing the gold seal in the split/splitless inlet, including removing the liner and column.
Guide for replacing the filter cartridge in the split vent line of the split/splitless inlet.
Step-by-step instructions for cleaning the split/splitless inlet, including removing parts and using solvent.
Procedure for baking out contaminants from the split/splitless inlet by setting temperatures and flows.
Lists parts and consumables for the purged packed inlet, including liners, adapters, and septa.
Diagram showing the exploded components of the purged packed inlet.
Step-by{0}by{0}step guide for installing a capillary column into the purged packed inlet.
Instructions for replacing the septum on the purged packed inlet, including safety precautions.
Procedure for cleaning the septum seat in the purged packed inlet assembly.
Details on how to install an adapter onto the purged packed inlet, including gathering materials.
Instructions for changing the O-ring on the purged packed inlet.
Procedure for changing the glass liner on the purged packed inlet.
Guide for installing an insulation cup on the purged packed inlet.
Step-by-step instructions for cleaning the purged packed inlet.
Procedure for baking out contaminants from the purged packed inlet.
Instructions for preparing and installing a packed metal column into the purged packed inlet.
Details on installing a packed column adapter onto a detector fitting.
Step-by-step guide for installing a packed glass column.
Procedure for conditioning a packed column according to manufacturer recommendations.
Instructions for installing ferrules onto a packed metal column.
Lists recommended parts for injections onto fused silica columns and recommended septa for the COC inlet.
Diagram showing the exploded components of the COC inlet.
Step-by-step guide for installing a capillary column into the COC inlet.
Procedure for checking the needle size compatibility with the column and insert for the COC inlet.
Instructions for replacing the septum on the COC inlet, with diagrams for different injection types.
Procedure for removing the existing insert and installing a new one in the COC inlet.
Step-by-step instructions for cleaning the COC inlet, including component removal and ultrasonic cleaning.
Details on installing the on-column needle support guide for the 7693A ALS injector.
Instructions for replacing the needle support assembly in a 7683B injector.
Step-by-step guide for replacing the needle in a syringe.
Procedure for replacing the fused silica needle in a syringe used with the COC inlet.
Procedure for baking out contaminants from the COC inlet by setting specific temperatures and flows.
Lists consumables and parts for the Multimode Inlet (MMI), including liners, nuts, and ferrules.
Diagram showing the exploded components of the Multimode Inlet (MMI).
Instructions for installing the column nut adapter onto the MMI inlet base.
Step-by-step guide for installing a capillary column into the Multimode Inlet (MMI).
Instructions for changing the septum on the Multimode Inlet (MMI).
Procedure for cleaning the septum seat in the insert assembly of the Multimode Inlet (MMI).
Details on how to change the liner and O-ring on the Multimode Inlet (MMI).
Guide for replacing the filter cartridge in the split vent line of the Multimode Inlet (MMI).
Step-by-step instructions for cleaning the Multimode Inlet (MMI), including liner removal and solvent cleaning.
Procedure for baking out contaminants from the Multimode Inlet (MMI).
Lists PTV liners, ferrules, syringes, and other consumables for the PTV inlet.
Diagram showing the exploded components of the PTV inlet.
Step-by-step guide for installing a capillary column into the PTV inlet.
Procedure for cleaning the septumless head assembly of the PTV inlet.
Instructions for replacing the PTFE ferrule on the septumless head of the PTV inlet.
Instructions for changing the septum on the PTV inlet.
Procedure for cleaning the septum seat in the septum head assembly of the PTV inlet.
Details on how to change the liner on the PTV inlet.
Instructions for replacing the inlet adapter for the PTV inlet.
Guide for replacing the filter cartridge in the split vent line of the PTV inlet.
Procedure for baking out contaminants from the PTV inlet.
Lists parts for the Volatiles Interface (VI), including clamping plates, adapters, nuts, and ferrules.
Diagram showing the exploded components of the Volatiles Interface (VI).
Step-by-step guide for installing a capillary column with the Volatiles Interface (VI).
Instructions for removing the Volatiles Interface (VI).
Procedure for cleaning the Volatiles Interface (VI) using an ultrasonic bath.
Details on how to install the Volatiles Interface (VI).
Guide for replacing the filter cartridge in the split vent line of the Volatiles Interface (VI).
Procedure for baking out contaminants from the VI inlet.
Lists nuts, ferrules, hardware, FID parts, subassemblies, and jets for the Flame Ionization Detector (FID).
Diagrams showing the exploded components of the Flame Ionization Detector (FID).
Guidance on selecting the appropriate FID jet based on application type and fitting.
Instructions for attaching a capillary column adapter to an adaptable FID.
Step-by-step guide for installing a capillary column into the Flame Ionization Detector (FID).
Procedure for replacing the collector assembly of the Flame Ionization Detector (FID).
Instructions for replacing the jet in the Flame Ionization Detector (FID).
Details on performing maintenance on the FID collector assembly, including gathering parts.
Procedure for checking the leakage current of the Flame Ionization Detector (FID).
Procedure for checking the baseline of the Flame Ionization Detector (FID).
Instructions for installing the insulation cup assembly on an adaptable FID.
Details on installing the optional PTFE chimney insert into the FID.
Procedure for baking out the FID detector.
Lists standard and optional TCD column hardware and consumables.
Step-by-step guide for installing a capillary column into the Thermal Conductivity Detector (TCD).
Instructions for installing an optional capillary column adapter for the Thermal Conductivity Detector (TCD).
Guide for installing a capillary column using the optional TCD capillary column adapter.
Procedure for baking out contaminants from the Thermal Conductivity Detector (TCD).
Provides crucial safety information regarding the uECD's radioactive isotope and handling precautions.
Lists consumables and parts for the micro-Electron Capture Detector (uECD).
Diagram showing the exploded components of the micro-Electron Capture Detector (uECD).
Instructions for replacing the liner and installing the makeup gas adapter for the uECD.
Step-by-step guide for installing a capillary column into the micro-Electron Capture Detector (uECD).
Procedure for installing the insulating cup for the micro-Electron Capture Detector (uECD).
Procedure for baking out contaminants from the micro-Electron Capture Detector (uECD).
Lists parts and consumables for the Nitrogen-Phosphorus Detector (NPD), including collectors, beads, and adapters.
Diagram showing the exploded components of the Nitrogen-Phosphorus Detector (NPD).
Guidance on selecting the appropriate NPD jet based on fitting type and application.
Instructions for attaching a capillary column adapter to an adaptable NPD.
Step-by-step guide for installing a capillary column into the Nitrogen-Phosphorus Detector (NPD).
Procedure for replacing the NPD bead assembly, including safety precautions and configuration.
Instructions for maintaining NPD collector, insulators, and jet, including replacement of related parts.
Procedure for checking the leakage current of the Nitrogen-Phosphorus Detector (NPD).
Procedure for baking out the Nitrogen-Phosphorus Detector (NPD).
Lists supplies for the Flame Photometric Detector (FPD+), including filters, ignitors, and adapters.
Diagrams showing the exploded components of the Flame Photometric Detector (FPD+).
Instructions for installing a packed column adapter in the Flame Photometric Detector (FPD+).
Step-by-step guide for attaching a capillary column to the Flame Photometric Detector (FPD+).
Procedure for changing the wavelength filter in the Flame Photometric Detector (FPD+).
Instructions for removing the cover of the Flame Photometric Detector (FPD+).
Procedure for replacing the ignitor in the Flame Photometric Detector (FPD+).
Instructions for installing the cover of the Flame Photometric Detector (FPD+).
Guidance on cleaning the brazement of the FPD+, noting precautions for the inert coating layer.
Lists consumables and parts for the Agilent G3435A/G3436A FPD detector.
Diagram showing the exploded components of the G3435A/G3436A FPD detector.
Instructions for installing a capillary column adapter into the FPD.
Step-by-step guide for attaching a capillary column to the FPD.
Procedure for changing the wavelength filter in the FPD detector, including PMT safety.
Instructions for removing the vent tube assembly from the FPD.
Procedure for replacing the ignitor in the FPD detector.
Instructions for installing the vent tube assembly and cover on the FPD.
Lists Aux EPC consumables and auxiliary channel frits, including O-rings and frits with flow characteristics.
Procedure for installing or replacing frits in the Auxiliary EPC block.
Lists consumables for the Pressure Control Module (PCM), including O-rings and proportional valves.
Procedure for calibrating the PCM interface, including zeroing the pressure sensor.
Instructions for installing or replacing frits in the Pressure Control Module (PCM).
Lists consumables and parts for valves, including gas sampling and liquid sampling valve types and loops.
Diagram showing the exploded components of GC rotary valves.
Instructions for replacing the sample loop in a gas sampling valve.
Procedure for aligning the rotary valve rotor, including adjusting the set screw.
Instructions for replacing a rotary valve within the valve box, including safety warnings.
Procedure for removing the upper valve box assembly.
Instructions for installing the upper valve box assembly, including coupling engagement.
Objective and steps for making leak-free tubing connections with Swagelok fittings.
Procedure for using a Swagelok Tee to supply gas from a single source to multiple inputs.