Maintaining the Split/Splitless Inlet 5
Maintaining Your GC 57
5
Grasp the liner with tweezers and pull it out.
6 Inspect the surface of the gold seal for graphite or rubber
septum contamination. If required, replace the gold seal.
(See “To Replace the Gold Seal on the Split/Splitless
Inlet” on page 59.)