58 Maintaining Your GC
5 Maintaining the Split/Splitless Inlet
Clean the inlet if there is visible or suspected
contamination. (See “To Clean the Split/Splitless Inlet” on
page 64.)
Clean O- ring residue from sealing surface.
7 Slide a new O- ring onto the replacement liner.
8 Return the liner to the inlet, pushing it all the way in
until the liner contacts the gold seal.
9 Line up the tab on the bottom of the septum assembly
with the slot on the insert assembly and push down to
connect. Slide the locking tab to the back.
10 Turn on the inlet. Allow the inlet and column to purge
with carrier gas for 15 minutes before heating the inlet or
the column oven.
11 Bakeout contaminants. (See “To Bakeout Contaminants
from the Split/Splitless Inlet” on page 66.)
12 Restore the analytical method.
13 Reset the liner counter. See To Reset an EMF Counter in
the Operation Manual.
14 If using the optional barcode scanner, scan the liner
information directly into your data system software
configuration, then download the changes to the GC.
15 Check for leaks.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals