Control Module Screens of the Well-plate Samplers 8
1100 Series WS MWS Reference Manual 251
Press ‘F1’ (Link Pump) to get access the screen, that allows you to choose one
of the pumps (if multiple pumps are configured), which will then be used for
the calculation of the flushout time of the sampler (depending on flow rate
and internal volume of the sampler).
By selecting More...Sample Flush-Out on the pull-down menu you can enter
the Sample Flush-Out Factor (specifies the (flush) volume of mobile phase
pumped through the injector after each injection before performing the
‘Automatic Delay Volume Reduction’ or the ‘Overlapped Injection’). Changes
must be acknowledged with F6 (Done).