Agilent InfinityLab LC Series Vialsamplers User Manual 110
3Using the Module
Using the Optional Integrated Column Compartment
The Plumbing button of the ChemStation Edit Columns table displays the
Column Assignment dialog box, which can also be displayed by selecting
Column Assignment from the context menu of the column compartment
Dashboard panel in the ChemStation's Method and Control view.
Figure 24 Column Assignment dialog box overview
The Column Assignment dialog box has three sections that give you
information about the column:
The Plumbing section shows the valve connections to the positions in the
column compartment.
NOTE
Make the connections to give the shortest distances between the valve ports and
the columns, and use a logical order (left column 1 to port 1-1, left column 2 to
port 2-2 and so on). Avoid leaving unused ports between used ones.