Agilent InfinityLab LC Series Vialsamplers User Manual 219
7Error Information  
Integrated Column Compartment (ICC) Heater Error Messages
Integrated Column Compartment (ICC) Heater 
Error Messages
Wait for temperature function timed out
Error ID: 30820
The system reported an error because reaching the set temperature took longer 
than expected. To verify the correct functioning of the heating, perform the 
Heater Test in the Agilent Lab Advisor software.
Heater over-current protection active
Error ID: 30821
The heating function of the Integrated Column Compartment is switched due to 
overcurrent.
To verify the correct functioning of the heating element, perform the Heater Test 
in the Agilent Lab Advisor software.
Probable cause Suggested actions
1
Connection problems
• Power off the system and verify that the ICC 
is correctly mounted in the sampler.
• Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Connection problems
• Power off the system and verify that the ICC 
is correctly mounted in the sampler.
• Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.