Agilent InfinityLab LC Series Vialsamplers User Manual 221
7Error Information  
Integrated Column Compartment (ICC) Heater Error Messages
Maximum heater temperature exceeded
Error ID: 30824
The heating function of the Integrated Column Compartment is switched off 
because the temperature registered by the temperature sensor exceeds the 
upper limit (100
 °C).
To verify the correct functioning of the heating element, perform the Heater Test 
in the Agilent Lab Advisor software
Column heater not connected
Error ID: 30825
The Integrated Column Compartment is either not connected or cannot be 
recognized by the system.
Probable cause Suggested actions
1
Connection problems
• Power off the system and verify that the ICC 
is correctly mounted in the sampler.
• Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Connection problems
• Power off the system and verify that the ICC 
is correctly mounted in the sampler.
• Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.