Agilent InfinityLab LC Series Vialsamplers User Manual 221
7Error Information
Integrated Column Compartment (ICC) Heater Error Messages
Maximum heater temperature exceeded
Error ID: 30824
The heating function of the Integrated Column Compartment is switched off
because the temperature registered by the temperature sensor exceeds the
upper limit (100
°C).
To verify the correct functioning of the heating element, perform the Heater Test
in the Agilent Lab Advisor software
Column heater not connected
Error ID: 30825
The Integrated Column Compartment is either not connected or cannot be
recognized by the system.
Probable cause Suggested actions
1
Connection problems
Power off the system and verify that the ICC
is correctly mounted in the sampler.
Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Connection problems
Power off the system and verify that the ICC
is correctly mounted in the sampler.
Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.